Erbe ICC 350 Electrosurgical unit

ERBE ICC 350 Electrosurgical Unit

Regular price
R 18,000.00
Sale price
R 18,000.00
Regular price
Sold out
Unit price
per 
Tax included.

This refurbished ERBE ICC 350 Electrosurgical Unit is the best model for monopolar and bipolar cutting and coagulation.

  • General surgery
  • Dermatology
  • Endoscopy
  • Gastroenterology
  • Vascular Surgery
  • Gynecology
  • Neurosurgery
  • Plastic Surgery
  • Oral surgery

Dimensions

  • Dimensions (W x H x D): 410 × 152 × 368 mm
  • Weight: 10.0 kg

Electrical

  • Rated power voltage: 240V/230V/115V/110V/100V± 10%
  • Rated power frequency: 50/60 Hz
  • Power current: 4.0 A at 230 – 240 V/8.0 A at 100-115 V
  • Power consumption in Standby mode: 25 watts
  • Power consumption at max. HF power: 620 watts 920 VA
  • Current consumption in Standby mode: 150 mA at 230 – 240V/300 mA at 100-115 V
  • Potential equalization terminal: Yes
  • Power fuses: 2 slow burn, 4 A at 230 – 240 V/8 A at 100-115 V

Cutting

  • HF power adjustment: With up/down increment keys
  • The form of HF voltage: Sine-shaped
  • The nominal frequency of HF voltage: 330 kHz
  • Cutting quality: 4 coagulation effects, selectable by key
  • The constancy of the coagulation effects: Automatically regulated
  • Support for initial incision and cutting: PPS, PowerPeakSystem
  • Monopolar AUTO CUT: With automatic voltage control
  • Monopolar HIGH CUT: with automatic arc control
  • Nominal HF output: 300 watts at 500 ohms
  • Peak HF power with PPS: 400 watts at 500 ohms
  • Monopolar ENDO CUT: 200 watts at max 200 ohms
  • BIPOLAR AUTO CUT: With automatic voltage control
  • Nominal HF output: 100 watts at 500 ohms

Coagulation

  • HF power adjustment: With up/down increment keys

Monopolar SOFT COAG

  • Form of HF current: Sine-shaped
  • Nominal frequency of HF voltage: 300 kHz
  • Nominal HF output: 120 watts at 125 ohms
  • End of coagulation: with Pedal or AUTO STOP

Monopolar FORCED COAG

  • Form of HF current: Pulse-modulated
  • Nominal frequency of HF voltage: 1 MHz
  • Nominal HF output: 120 atts at 350 ohms

Monopolar SPRAY COAG

  • Form of HF current: Pulse-modulated
  • Nominal frequency of HF voltage: 1 MHz
  • Nominal HF output: 120 watts at 500 ohms

Biopolar COAG

  • Form of HF current: Sine-shaped
  • The nominal frequency of HF voltage: 330 kHz
  • Nominal HF output: 120 watts at 125 ohms
  • Start of coagulation: By foot pedal or AUTOSTART
  • End of coagulation: By foot pedal or AUTO STOP